ACUSON SC2000 PRIME Ultrasound System - Mainframe (2018-01-31)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Siemens Medical Solutions USA, Inc.
685 Middlefield Road
Mountain View
94043
California
UNITED STATES
ACUSON SC2000 PRIME Ultrasound System - Mainframe10433816

Introduction

Affected products

ACUSON SC2000 PRIME Ultrasound System - Mainframe

Reason

Siemens is notifying users of a potential issue with the ACUSON SC2000 ultrasound system with software versions VB10A, VB10B, VB10D and VB10E. The 3D eSie PISA volume analysis application quantifies valvular regurgitation. While imaging with the Z6Ms volume transesophageal echocardiography transducer, this application may underestimate the EROA (Effective Regurgitant Orifice Area) in comparison to the same patient results obtained with the 4Z1c volume transthoracic echocardiography transducer.