Affected products
PENTAX Video Duodenoscope
In March 2017, as part of field action 2017-001-c, Pentax Medical informed customers about a potential cracks and gaps in the distal cap of the ED-3490TK, and instituted a distal cap inspection program. During discussion with the US FDA as part of this field action, Pentax Medical agreed to redesign the ED-3490TK's forceps elevator mechanism, O-Ring, and Distal End to reduce the risk of cross-contamination resulting from body fluids entering portions of the ED-3490TK that cannot be easily cleaned.