Introduction

Affected products

AlterG M320

Reason

A condition was reported by a distributor in which an M320 unit unexpectedly surged in speed to the maximum allowed by the treadmill. The treadmill e-stop was not effective. Failure traced to a specific diode on the lower motor controller.