Health Products

Osscora Surgical Set (2018-02-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
E01039
Manufacturer
Cochlear Bone Anchored Solutions Ab
Konstruktionsvagen 14
Molnlycke
43533
SWEDEN
Osscora Surgical Set91053

Introduction

Affected products

Osscora Surgical Set

Reason

The supplier W&H Dentalwerk (in Burmoos, Austria) notified Cochlear Bone Anchored Solutions (CBAS) of a potential malfunction in the foot control included in Osscora Surgical Set (part# 91053). Foot control may not switch back to zero position when pedal is released-causing motor to continue rotation. Drilling mode not affected. Automatic torque stop is fully functional.