Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Bovie Medical Corporation
5115 Ulmerton Road
Clearwater
33760
Florida
UNITED STATES
J-PLASMA HANDPIECE, BLADE TIP, PISTOL GRIPBVX-150B

Introduction

Affected products

J-PLASMA HANDPIECE, BLADE TIP, PISTOL GRIP

Reason

Bovie Medical has become aware of an issue caused by incomplete insertion of the J-Plasma Handpiece cable plug into the generator receptacle. When the J-Plasma Handpiece with generator is used during surgical procedures without the cable plug being completely inserted into the generator, the green indicator light to the left of the receptacle will illuminate event with an incomplete cable plug insertion. There have been no patient injuries reported as a result of this issue.