Somatom Emotion 16-Slice Configuration and Somatom Perspective - CT Scanner (2018-02-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
121001
Manufacturer
Siemens Shanghai Medical Equipment Ltd.
278 Zhouzhu Road
B. SOMATOM PERSPECTIVE - CT SCANNER10495568
85828
Manufacturer
Siemens Shanghai Medical Equipment Ltd.
278 Zhouzhu Road
A. SOMATOM EMOTION 16-SLICE CONFIGURATION10165877

Introduction

Affected products
  1. SOMATOM EMOTION 16-SLICE CONFIGURATION
  2. SOMATOM PERSPECTIVE - CT SCANNER

Reason

Siemens healthcare became aware of possible incorrect tube current calculations by the Care Dose4d algorithm for head scans based on p.a. (posterior-anterior) or a.p. topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure.