MYLAB 70 X Vision and XVG, MYLAB TWICE and CLASS C (2018-01-22) Jan 22, 2018 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number Not applicable Manufacturer Esaote S.P.A Via Enrico Melen 77 Genova 16152 ITALY D. MYLAB CLASS C 101625000 Not applicable Manufacturer Esaote S.P.A Via Enrico Melen 77 Genova 16152 ITALY B. MYLAB 70 XVG 9806100000 Not applicable Manufacturer Esaote S.P.A Via Enrico Melen 77 Genova 16152 ITALY C. MYLAB TWICE 101620000 Not applicable Manufacturer Esaote S.P.A Via Enrico Melen 77 Genova 16152 ITALY A. MYLAB 70 X VISION 9600169000
Affected products
MYLAB 70 X VISION
MYLAB 70 XVG
MYLAB TWICE
MYLAB CLASS C
Probes default settings in Elaxto Mode (Elastosonography Application) are found to be not appropriate for certain probes, resulting in overheating of the probe head.