Brilliance CT Big Bore System-Main Unit, Brilliance ICT System - System, Brilliance ICT SP System - Main and Ingenuity CT System (2018-02-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 Miner Road
Cleveland
44143
UNITED STATES
D. INGENUITY CT SYSTEM728323
Not applicable
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 Miner Road
Cleveland
44143
UNITED STATES
B. BRILLIANCE ICT SYSTEM- SYSTEM728306
Not applicable
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 Miner Road
Cleveland
44143
UNITED STATES
C. BRILLIANCE ICT SP SYSTEM - MAIN728311
Not applicable
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 Miner Road
Cleveland
44143
UNITED STATES
A. BRILLIANCE CT BIG BORE SYSTEM-MAIN UNIT4535 670 88051

Introduction

Affected products
  1. BRILLIANCE CT BIG BORE SYSTEM-MAIN UNIT
  2. BRILLIANCE ICT SYSTEM- SYSTEM
  3. BRILLIANCE ICT SP SYSTEM - MAIN
  4. INGENUITY CT SYSTEM

Reason

Upon closing the Emergency Stop (E-STOP) the system performs an internal check. As a result of the motion controller in the couch, there is a delay in the response to the internal check. This delay may cause the couch carbon top to enter a free float state, which may allow it to move during patient load or unload activities. This free float motion results in the ESTOP being opened due to the unexpected couch motion. This issue occurs frequently after closing E-STOP. The motion controller in the couch detects a power problem and disables the horizontal motor, which allows the couch top to free-float*.                                         

*The term free float refers to the couch top being in a state that is not locked in position by the brake or being moved in a controlled fashion manual or automatic motion.