HemosIL Factor XII Deficient Plasma are not meeting their labeled on-board instrument stability claim for the ACL Top Family and ACL Top Family 50 Series of 24 hours at 15 C, the affected lots numbers are only stable for 2 hours on-board the instrument. Consequently, there is a potential risk of falsely lowered patient results if this reagent is maintain greater than 2 hours on-board the instrument.