Lot or serial numberCompaniesBrand nameModel or catalog number
2125747
Manufacturer
Oxoid Limited
Wade Road
Basingstoke, Hampshire
RG24 8PW
UNITED KINGDOM
BRILLIANCE SALMONELLA AGARCM1092B

Introduction

Affected products

BRILLIANCE SALMONELLA AGAR

Reason

The lot number 2125747 is failing to meet the inhibition specification for the Quality Control organism Escherichia coli ATCC 2592 i.e. 2 log inhibition. This lot however supports the growth of Salmonella spp. to the required specification and not all E.coli species are impacted and acceptable inhibition was observed with some strains.