Health Products

TDWorkstation (2018-02-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
DR35041
Manufacturer
TECHNIDATA S.A.S.
387 Avenue Jean Kuntzmann
Montbonnot-Saint-Martin
38330
FRANCE
TDWORKSTATIONTD-WST-01

Introduction

Affected products

TDWORKSTATION

Reason

Software anomaly (reference dr35041) with TD-Synergy version v11.31 or later. This anomaly results in consequences for users only when the TD-Synergy system is connected to another system and uses one of these two communication methods. Transmission of results in HL7 format from TD-Synergy to a host system including results of date type. Sharing of TD-Synergy data with a reference laboratory including complementary parameters of date type.

This anomaly affects only results of date type or complementary parameters of date type.

Nature of the defect (real or potential):
  • Depending on the date format used, the system may invert the date’s day and month. For example, the system can transmit a date as “February 3, 2015” instead of “March 2, 2015”
  • The system transmits results of date type or review requests with complementary parameters of date type in HL7 format.


The communication module’s service window uses a local account system that has a regional parameter corresponding to common wording that uses a date format in which the day precedes the month (e.g., “day/month/year” format such as “dd/mm/yy” or “d/m/yy”).
Or
The communication module’s service window uses a specific user account, and this user uses a date format in which the day precedes the month (e.g., “day/month/year” format such as “dd/mm/yy” or “d/m/yy”).

The defect does not occur if the date format used is “month/day/year” (mm/dd/yyyy).