VITEK 2 Companct 30 System - Instrument (2018-02-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Biomerieux Inc.
100 Rudolphe Street
Durham
27712
North Carolina
UNITED STATES
VITEK 2 COMPACT 30 SYSTEM - INSTRUMENT27530

Introduction

Affected products

VITEK 2 COMPACT 30 SYSTEM - INSTRUMENT

Reason

Internal investigation has demonstrated that in specific cases when a card is processed in, and ejected from, a slot number greater than 15 or 30 (all VITEK 2 Compact Reader Carousels contain 60 physical slots), respectively, the final data packet for the card fails to be accepted by the computer, and the isolate status remains at "preliminary". When the anomaly occurs, subsequent data for current and future cards will not be transferred to the computer until the offending data packet is removed. Since the card has completed processing, as evidenced by the card ejection from the VITEK 2 compact reader, the results indicated in the lab report are the final results. As the card data for the ejected card will be lost after record removal, it is recommended the user print the lab report prior to removal of the record.