Health Products

Biograph MMR System (2018-01-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
151014
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
Biograph MMR System10433372

Introduction

Affected products

Biograph MMR System

Reason

Siemens has identified a mismatch between orientation labels and displayed anatomy (e.g. in the 3D task card) within the software version syngo MR E11P for Biograph mMR. This will lead to a wrong attenuation correction of the pet data.