Health Products

Graphic Cases for 6.5 mm and 7.3 mm Cannulated Screw Sets (2018-03-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
FL00860Graphic Cases for 6.5 mm and 7.3 mm Cannulated Screw Sets
Manufacturer
Synthes USA LLC
1101 SYNTHES AVENUE
Monument
80132
Colorado
UNITED STATES
690.375-US

Introduction

Affected products

Graphic Cases for 6.5 mm and 7.3 mm Cannulated Screw Sets

Reason

DePuy Synthes is conducting a lot-specific recall of the graphic cases for 6.5mm and 7.3mm cannulated screw sets because they contain screw gauges that incorrectly measure screws 5.41mm shorter than the actual overall screw length. Thus, when using the measuring gauge, the surgeon would potentially be inserting a screw which is 5.41mm shorter than intended.