The internal mains wiring does not fully comply with the requirements of the medical electrical equipment standard IEC60601-1 Edition 3.0. IEC60601-1 states that conductors and connectors of medical electrical equipment shall be secured or insulated so that accidental detachment shall not result in a hazardous situation. The K-MAR-5200 mains wiring is secured, but the mechanism is not considered suitable under the current version of IEC60601-1.