iQ200 Sprint and Elite Automated Urine Microscopy Analyzer, iQ200 Select Urinanlysis System - Analyzer, iQ200 Elite Plus and Pro, iQ Select Plus and Pro, iQ Sprint Plus and Pro (2018-02-26)
C. IQ 200 ELITE AUTOMATED URINE MICROSCOPY ANALYZER
800-3920
All lots.
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES
IQ 200 SPRINT AUTOMATED URINE MICROSCOPY ANALYZER
700-3322
Introduction
Affected products
IQ 200 SPRINT AUTOMATED URINE MICROSCOPY ANALYZER
IQ200 SELECT URINALYSIS SYSTEM - ANALYZER
IQ 200 ELITE AUTOMATED URINE MICROSCOPY ANALYZER
IQ 200 ELITE PLUS
IQ 200 ELITE PRO
IQ 200 SELECT PRO
IQ 200 SELECT PLUS
IQ 200 SPRINT PRO
IQ 200 SPRINT PLUS
Reason
Beckman coulter has determined that there is a potential for under-reporting casts. This can occur if per High-Power Field (HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal threshold and/or grading format is set up based on reporting "per Low-Power Field (LPF)" or “per microliter (ul)”. This may occur during initial method validation or if settings are altered after the initial validation.