IH-1000 (2018-03-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Diamed GMBH
Pra Rond 23
Cressier FR
1785
SWITZERLAND
IH-1000001000

Introduction

Affected products

IH-1000

Reason

When a reagent vial with an unreadable barcode is loaded on a reagents rack on IH-1000 (software version 04.07.02) and a test is performed, despite the fact the first reagent is correctly pipetted, the instrument will pipette in the vial with unreadable barcode for all the remaining wells.