SERVO-i Ventilator System (2018-03-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
83775 - 87140
Manufacturer
Maquet Critical Care AB
Rontgenvagen 2
Solna
17154
SWEDEN
SERVO-i Ventilator System64 87 800 E407E

Introduction

Affected products

SERVO-i Ventilator System

Reason

Under certain conditions, an internal communication error might prevent the safety valve to close automatically. As a consequence, ventilation will stop a high priority alarm, as well as clinical alarms, will be triggered. We have no reports of any permanent injuries. This issue will stop ventilation, the safety valve open, thus allowing the patient to breathe freely, but without ventilator support. The high priority technical error alarm TE 11 will be activated simultaneously with other clinical alarms.