Achieva and Intera 1.5T MRI System - Main Unit (2018-03-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
Intera 1.5T MRI System-Main UnitINTERA 1.5T
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
Achieva 1.5T MRI System- Main UnitACHIEVA 1.5T

Introduction

Affected products

Reason

In rare cases, helium gas may escape into the examination room when superconductivity is lost (also known as quenching of the magnet) and at the same time the designed helium ventilation path is impeded. Philips has become aware of one instance where this happened in a Philips system. Investigation found the metal burst disk assembly in the vent path did not perform according to specifications.