EMS Power-LOAD (2018-03-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
Stryker Medical
3800 E Centre Avenue
Portage
49002
UNITED STATES
EMS Power-LOAD6390-000-000

Introduction

Affected products

EMS Power-LOAD

Reason

It was reported via three complaints that after installation of the power-LOAD, the floor plate attachment bracket (PN: 6390-101-108) was found to be broken.
After installation of the power-LOAD, the floor plate attachment brackets (PN: 6390-101-108) which allow the power-LOAD anchor to be secured to the power-load floor plate may be found to be broken after removing the anchor. All power-LOAD units manufactured from January 1, 2016 to December 31, 2016, may potentially exhibit this issue.