Infinity M540 (2018-03-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
Draeger Medical Systems, Inc.
3135 Quarry Road
Telford
18969
UNITED STATES
Infinity M540MS20401

Introduction

Affected products

Infinity M540

Reason

Draeger learned about an issue with the infinity 540 patient monitor with software versions vg4.1 and below. When the m540 patient monitor is used with and connected to the Infinity MCable Mainstream CO2 (m11.1) sensor revision 16, the m540 patient monitor will reboot if the sensor enters a reduced accuracy mode. If the m540 patient monitor reboots 3 times within a 10 minute period or less, the m540 patient monitor will go enter a fail-state, resulting in a resetting to factory defaults and loss of patient settings and stored patient data. Revision 16 sensors were introduced in mid-2017. Use of revision 15 and earlier sensors will not cause this issue to occur.