Health Products

Contact Matrix Plus and Contact Ring Refill (2018-03-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
L03AQ
Manufacturer
Danville Materials
2875 Loker Avenue East
Carlsbad
92010
California
UNITED STATES
B. Contact Matrix Plus89439
I264R
Manufacturer
Danville Materials
2875 Loker Avenue East
Carlsbad
92010
California
UNITED STATES
Contact Matrix Contact Ring Refill88038

Introduction

Affected products
  1. Contact Matrix Contact Ring Refill
  2. Contact Matrix Plus

Reason

In the affected product, the mesial/distal shape of the inward contact rings is potentially incorrect.  Specifically, the inward contact rings should have a distal (narrowing) tine shape when placed with the ring oriented toward the front of the mouth.  Some contact rings were found to have (widening) tines.