Health Products

Cobas e801 Syringes (2018-03-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1601-01 To 18E6-10
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
Cobas e801 Syringes07682913001

Introduction

Affected products

Cobas e801 Syringes

Reason

The plunger of the Cobas e801 syringe assembly may be mounted in a tilted position which applies mechanical stress to the plunger and may cause r1, r2 or prewash liquids to be pipetted incorrectly.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime