Dimension Vista System - Multi 1 Sample Diluent (MULTI 1 SDIL) and Multi 2 Sample Diluent (MULTI 2 SDIL), CTNI Sample Diluent (CITNI SDIL) (2018-03-26)
B. DIMENSION VISTA SYSTEM MULTI 1 SAMPLE DILUENT (MULTI 1 SDIL)
10469971
7BDA34
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
C. CTNI SAMPLE DILUENT (CTNI SDIL)
10445205
7CDA79
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
A. DIMENSION VISTA SYSTEM - MULTI 2 SAMPLE DILUENT (MULTI 2 SDIL)
10483586
Introduction
Affected products
A. DIMENSION VISTA SYSTEM - MULTI 2 SAMPLE DILUENT (MULTI 2 SDIL)
B. DIMENSION VISTA SYSTEM MULTI 1 SAMPLE DILUENT (MULTI 1 SDIL)
C. CTNI SAMPLE DILUENT (CTNI SDIL)
Reason
Siemens Healthcare Diagnostics has confirmed certain Dimension Vista diluent lots may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is auto-diluted onboard the Dimension Vista System, due to a reduced volume of diluent being pipetted from the vial. Only a portion of each lot has defective septums and only a portion of visually defective septums are functionally defective.