Health Products

HeartMate III LVAS Implant Kit (2018-04-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Thoratec Corporation
6035 STONERIDGE DRIVE
Pleasanton
94588
California
UNITED STATES
HeartMate III LVAS Implant Kit106524US

Introduction

Affected products

HeartMate III LVAS Implant Kit

Reason

Abbott is advising our physician partners that we have received reports of outflow graft twist occlusions in the HeartMate 3 (HM3) Left Ventricular Assist System. As a result, patients whose devices experience these outflow graft occlusions will experience a persistent low flow alarm. Outflow graft twists can result in serious adverse events such as hemodynamic compromise, thrombus, and death.