ADVIA Chemistry System - Urea Nitrogen (UN) Assay (2018-04-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
408143
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
ADVIA Chemistry System - Urea Nitrogen (UN) Assay03040257

Introduction

Affected products

ADVIA Chemistry System - Urea Nitrogen (UN) Assay

Reason

Siemens Healthcare Diagnostics has confirmed that the Urea Nitrogen Reagent Kit Lots 408143 and 418290 for use on ADVIA 1800, 2400 and XPT Chemistry Systems demonstrate an increased incidence of calibration failure. Results cannot be generated by the system when calibration fails. If calibration passes, there is a potential for falsely elevated patient and quality control results. Internal testing has also shown that results have the potential to be falsely increased by approximately 10 mg/dl (3.57 mmol/l) across the analytical range. Quality controls may not always detect the elevated results.