Health Products

Sterile C Section Pack, Bone Marrow Tray, Sterile Labour and Delivery Pack, Sterile Obstetrical Pack, Sterile Delivery Pack, and Bone Marrow Biopsy Pack from Cardinal Health (2018-04-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
Cardinal Health
1500 Waukegan Road
Waukegan
60085
Illinois
UNITED STATES
F. BONE MARROW BIOPSY PACKSBA30BMHMA
More than 100 numbers, contact manufacturer.
Manufacturer
Cardinal Health
1500 Waukegan Road
Waukegan
60085
Illinois
UNITED STATES
D. STERILE OBSTETRICAL PACKSMA30OBPNG
More than 100 numbers, contact manufacturer.
Manufacturer
Cardinal Health
1500 Waukegan Road
Waukegan
60085
Illinois
UNITED STATES
E. STERILE DELIVERY PACKSMA30DLCLG
More than 100 numbers, contact manufacturer.
Manufacturer
Cardinal Health
1500 Waukegan Road
Waukegan
60085
Illinois
UNITED STATES
B. BONE MARROW TRAYSOP30BM00C
More than 100 numbers, contact manufacturer.
Manufacturer
Cardinal Health
1500 Waukegan Road
Waukegan
60085
Illinois
UNITED STATES
C. STERILE LABOUR AND DELIVERY PACKSMA30BILBJ
More than 100 numbers, contact manufacturer.
Manufacturer
Cardinal Health
1500 Waukegan Road
Waukegan
60085
Illinois
UNITED STATES
A. STERILE C SECTION PACKSMA30CSCLG

Introduction

Affected products

A. STERILE C SECTION PACK
B. BONE MARROW TRAY
C. STERILE LABOUR AND DELIVERY PACK
D. STERILE OBSTETRICAL PACK
E. STERILE DELIVERY PACK
F. BONE MARROW BIOPSY PACK

Reason

Cardinal Health has been advised of an advisory notice on BD Vacutainer Top Tubes contained in these kits.  BD has recently determined that the curing agent thiuram, used in the processing of rubber stoppers since 1996, can lead to release of Carbon Disulfide (CS2) and Carbonyl Sulfide (COS) gases into the headspace of the tubes. These gases can dissolve into the blood sample to chelate lead, and therefore interfere with the accuracy of the Anodic Stripping Voltammetry (ASV) testing methodology. Although the thiuram has been used since at least 1996, the release of CS2 and COS gases into the headspace and their effect on ASV methodology was only recently observed. ASV is the methodology used in magellan diagnostics' leadcare testing systems.