Health Products

Various Medlines' Custom Procedure Packs (2018-04-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
17YB1732F. Vaginal Delivery Pack - LF
Manufacturer
Medline Industries Inc.
Three Lakes Drive
Northfield
60093
Illinois
UNITED STATES
DYNJ45914X
17XB2402D. New Tissue Retrieval MH
Manufacturer
Medline Industries Inc.
Three Lakes Drive
Northfield
60093
Illinois
UNITED STATES
DYNJ903849X
18BB9026E. Ensemble A Cesarienne-LF
Manufacturer
Medline Industries Inc.
Three Lakes Drive
Northfield
60093
Illinois
UNITED STATES
DYNJ57105X
17KB1919B. Vaginal Delivery Pack
Manufacturer
Medline Industries Inc.
Three Lakes Drive
Northfield
60093
Illinois
UNITED STATES
DYNJ47920X
18MB5551C. C-Section Pack-LF / C-Section-LF
Manufacturer
Medline Industries Inc.
Three Lakes Drive
Northfield
60093
Illinois
UNITED STATES
DYNJ45915X
17WB9664A. Cesarienne-LF / C Section BJC-LF
Manufacturer
Medline Industries Inc.
Three Lakes Drive
Northfield
60093
Illinois
UNITED STATES
DYNJ41672X

Introduction

Affected products

A. Cesarienne-LF / C Section BJC-LF
B. Vaginal Delivery Pack
C. C-Section Pack-LF / C-Section-LF
D. New Tissue Retrieval MH
E. Ensemble A Cesarienne-LF
F. Vaginal Delivery Pack - LF

Reason

Medline Industries received a Medical Device Correction Notice from BD concerning their BD Vacutainer EDTA Lavender, Tan & Pink Top Tubes & BD Vacutainer Lithium Heparin Green Top Tubes. BD has recently determined that the curing agent thiuram, used in the processing of rubber stoppers since 1996, can lead to release of carbon disulfide (CS2) and carbonyl sulfide (COS) gases into the headspace of the tubes. These gases can dissolve into the blood sample to chelate lead, and therefore interfere with the accuracy of the Anodic Stripping Voltammetry (ASV) testing methodology.
The BD Vacutainer EDTA Lavender, Tan and Pink Top Tubes & BD Vacutainer Lithium Heparin Green Top Tubes are a component in several Medline Custom Procedure Packs.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime