Biograph mMR System (2018-04-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
151014
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
Biograph mMR System10433372

Introduction

Affected products

Biograph mMR System

Reason

Siemens has identified the potential for a mismatch between PET and MR images in z-direction when viewing fused images on select mMR systems. The issue is visible when users are viewing fused PET and MR images e.g. in the syngo 3D Fusion task card. When PET and MR images are compared thoroughly in an overview layout (side-by-side) e.g. in syngo.via, the effect can be observed, too.

The original MR and the non-attenuation-corrected PET images are presented correctly, but due to a miscalculation the alignment of the fused MR and PET images, both non-attenuation-corrected (NAC) and attenuation-corrected (AC) is not performed as expected. Also due to the same miscalculation, the quality of PET AC images may be degraded slightly.