Health Products

T2 Ankle Arthrodesis Nail, IEFT, Non-Sterile, T2 Femur System - Femoral Nails, T2 Tibial Nail System - Nail Standard (Sterile) and T2 Recon Nail System - Titanium (2018-04-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
K07F870D. T2 Recon Nail system - Titanium
Manufacturer
Stryker Trauma GMBH
Prof.-Kuntscher-Str.1-5
Schoenkirchen
24232
GERMANY
1844-0940S
K078C77B. T2 Femur System - Femoral Nails
Manufacturer
Stryker Trauma GMBH
Prof.-Kuntscher-Str.1-5
Schoenkirchen
24232
GERMANY
1825-1144S
K074846C. T2 Tibial Nail System - Nail Standard (Sterile)
Manufacturer
Stryker Trauma GMBH
Prof.-Kuntscher-Str.1-5
Schoenkirchen
24232
GERMANY
1822-1331S
K07AFEAA. T2 Ankle Arthrodesis Nail, IEFT, Non-Sterile
Manufacturer
Stryker Trauma GMBH
Prof.-Kuntscher-Str.1-5
Schoenkirchen
24232
GERMANY
1818-1130

Introduction

Affected products

A. T2 Ankle Arthrodesis Nail, IEFT, Non-Sterile
B. T2 Femur System - Femoral Nails
C. T2 Tibial Nail System - Nail Standard (Sterile)
D. T2 Recon Nail system - Titanium

Reason

On April 9, 2018, the manufacturer discovered that potentially out-of-specification products may have left the factory. To date, there have been no customer complaints logged for this issue. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime