Health Products

Elecsys Vitamin D total II (2018-04-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.B. VITAMIN D TOTAL II
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
07464215190
All lots.A. ELECSYS VITAMIN D TOTAL II
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
07028148190

Introduction

Affected products

A. ELECSYS VITAMIN D TOTAL II
B. VITAMIN D TOTAL II

Reason

During the implementation of the Elecsys Vitamin D total Gen. II Assay, Customers reported non-reproducible, false high results. These discrepant results were observed during assay validation of Elecsys Vitamin D total II where samples were run in duplicate and, during method comparisons with Elecsys Vitamin D total. The issue manifests as follows: The first result is elevated, either above the upper end of the measuring range (>100 ng/ml or >250 nmol/ml) or within the measuring range, and repeats are significantly lower.