Health Products

da Vinci Xi EndoWrist Stapler 45 Reload (2018-04-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
M10161214
Manufacturer
Intuitive Surgical Inc.
1266 Kifer Road, Building 100, 101, 102, 103
Sunnyvale
94086
UNITED STATES
da Vinci Xi EndoWrist Stapler 45 Reload48445G

Introduction

Affected products

da Vinci Xi EndoWrist Stapler 45 Reload

Reason

Specific lots of the da Vinci Xi Stapler 45 Blue and Green Reloads may be missing a component of the knife mechanism. Without this component, the reload knife will not move and cut tissue when the stapler is fired. As a result, the user may see a staple line that has all the staples formed, but no tissue transection.  The only manner by which a user will be made aware of the issue is through visual inspection of the tissue after using the stapler.