Affected products
A. Heartstart XL
B. HEARTSTART MRX Defibrillator/Monitor
C. HEARTSTART XL+
When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference from
electrical fast transients (EFTs):
- On the HeartStart MRx and HeartStart XL, EFT noise may be misinterpreted as an R-wave.
- On the HeartStart XL+, EFT noise can disable ECG monitoring and potentially interrupt demand mode pacing*.
*Note: It is contrary to the XL+ Instructions for Use to perform demand mode pacing while using the ECG out cable or obtaining the ECG signal from a bedside monitor. The XL+ Instructions for Use includes the following warning: "When pacing in Demand Mode, the ECG cable from the patient must be directly connected to the HeartStart XL+." If the user follows this warning, this problem cannot occur on the XL+.