Health Products

Philips OmniDiagnost and OmniDiagnost Eleva (2018-05-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
B. OMNIDIAGNOST ELEVA708027
Not applicable
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
A. PHILIPS OMNIDIAGNOSTOMNIDIAGNOST

Introduction

Affected products

A. PHILIPS OMNIDIAGNOST
B. OMNIDIAGNOST ELEVA

Reason

Philips has been reported instances in which the table of the OmniDiagnost system (OD-Eleva) suddenly started to rotate from 0 to 90 degree (table arm down) with high speed. The cause of this unexpected rotation movement is related to an issue with the fixation of the upper and lower tilt actuator. If the system rotates uncontrolled, there is a risk of injury for the patient on the table. A rotating table also might hit the user or bystander standing near the system.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime