Health Products

HistoCore PELORIS 3 (2018-05-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
45110009
Manufacturer
Leica Biosystems Melbourne Pty. Ltd.
495 Blackburn Road
Mount Waverley, Victoria
3149
AUSTRALIA
HistoCore PELORIS 345.0005

Introduction

Affected products

HistoCore PELORIS 3

Reason

If operator initiates an operation in second retort or the protocol running concurrently in second retort causes a dialog box to appear when draining first retort at completion of protocol, a software defect may interrupt instrument workflow sequences in the first retort resulting in the impeller continuing to rotate.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime