Heartware Ventricular Assist System- Controller/ Controller Kits, AC Adapter, DC Adapter, and Battery Pack (2018-05-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Heartware Inc.
14400 NW 60th Avenue
Miami Lakes
33014
Florida
UNITED STATES
D. Heartware Ventricular Assist System- Battery Pack1650DE
All lots.
Manufacturer
Heartware Inc.
14400 NW 60th Avenue
Miami Lakes
33014
Florida
UNITED STATES
B. Heartware Ventricular Assist System- AC Adapter1430CA
All lots.
Manufacturer
Heartware Inc.
14400 NW 60th Avenue
Miami Lakes
33014
Florida
UNITED STATES
C. Heartware Ventricular Assist System- DC Adapter1440
All lots.
Manufacturer
Heartware Inc.
14400 NW 60th Avenue
Miami Lakes
33014
Florida
UNITED STATES
A. Heartware Ventricular Assist System- Controller/ Controller Kits1407CA

Introduction

Affected products

A. Heartware Ventricular Assist System- Controller/ Controller Kits
B. Heartware Ventricular Assist System- AC Adapter
C. Heartware Ventricular Assist System- DC Adapter
D. Heartware Ventricular Assist System- Battery Pack

Reason

Manufacturer is providing this notification to inform customers of the potential for a transient interruption of the electrical connection between an HVAD system power source (battery, ac adapter, or dc adapter) and the HVAD controller that may result in unintended power switching to the secondary power source and/or unexpected audible tones ("beeping"). This interruption, which occurs while the power source remains physically connected, is due to oxidation of connecting surfaces between a power source connector and the controller's power source socket, and typically lasts 1-2 seconds.
Unexpected beeping occurs when the interruption automatically resolves and may cause confusion to the patient or caregiver, as the controller may display sufficient battery capacity or AC/DC connectivity at the time of the audible tone. A critical battery alarm may also be momentarily displayed due to this phenomenon.