VeriQ and VeriQ C System (2018-05-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Medistim ASA
Økernveien 94, Oslo
0579
NORWAY
B. VeriQ C SystemVQ4122C
All lots.
Manufacturer
Medistim ASA
Økernveien 94, Oslo
0579
NORWAY
A. VeriQ SystemVQ4122

Introduction

Affected products

A. VeriQ System
B. VeriQ C System

Reason

This communication from Medistim was issued due to the possibility that flow measurement channels on Medistim systems can operate with a significant zero-point offset value, caused by Electrostatic Discharge (ESD). Medistim has therefore further issued updated Instructions For Use (IFU) which clarifies and emphasizes the importance of routinely performing the probe and system functionality test prior to use in order to ensure that the system and probes function appropriately. There is no product retrieval associated with this field safety notice. This is a notification only, pertaining to the updated IFU.