Ortho confirmed customer complaints regarding the potential for imprecise results when using specific lots of the VITROS PHYT Slides.
Ortho performed internal testing to confirm the issue using several lots of coating 0165. The results obtained were either positively or negatively biased across the VITROS PHYT Slides measuring range of 3.00-40.00 ug/ml (11.88-158.40 umol/l).