Health Products

VITROS Chemistry Products PHYT Slides (2018-05-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.VITROS Chemistry Products PHYT Slides
Manufacturer
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
8298671

Introduction

Affected products

VITROS Chemistry Products PHYT Slides

Reason

Ortho confirmed customer complaints regarding the potential for imprecise results when using specific lots of the VITROS PHYT Slides. Ortho performed internal testing to confirm the issue using several lots of coating 0165. The results obtained were either positively or negatively biased across the VITROS PHYT Slides measuring range of 3.00-40.00 ug/ml (11.88-158.40 umol/l).