Dimension Vista System - A1-Antitrypsin Assay (2018-05-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
17279MADimension Vista System -A1-Antitrypsin Assay
Manufacturer
Siemens Healthcare Diagnostics Products Gmbh
Emil-Von-Behring-Str. 76
Marburg
35041
GERMANY
K7014

Introduction

Affected products

Dimension Vista System -A1-Antitrypsin Assay

Reason

Siemens is notifying customers to advise of a high rate of calibration failure for Dimension Vista Flex reagent cartridge A1AT lot 17279MA. The calibration failure is due to a Mean Absolute Relative Deviation (MARD = Percent Deviation) > 5% for the instrument's automatic calibration acceptance criterion. Siemens Healthcare Diagnostics Products GmbH is currently investigating this issue.