ABBOTT PRISM HIV O Plus Assay Kit (2018-05-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
83006M500
Manufacturer
Abbott Laboratories Diagnostics Division
100 ABBOTT PARK ROAD
ABBOTT PARK
60064
Illinois
UNITED STATES
ABBOTT PRISM HIV O Plus Assay Kit3L68-68

Introduction

Affected products

ABBOTT PRISM HIV O Plus Assay Kit

Reason

Abbott has received customer complaints for calibration failures associated with PRISM HIV O Plus assay kit (list number 3l68-68) due to the group 0 positive assay control (opc) exceeding the prism software sample to cut-off (S/CO) upper specification. When the opc exceeds the upper specification limit, the calibration report notes an assay control failure and no results are reported for the run. An investigation to determine the cause of this issue is in progress.