The Method Sheet of Elecsys AMH was updated with respect to the Biotin interference declaration in the unit nmol/L. Previously the claim stated incorrectly less or equal to 143 nmol/L; this was corrected to less or equal to 123 nmol/L. The root cause of this issue was erroneous value transmission to labeling.
Important: The declared value in unit ng/mL was correct at 30 ng/mL and does not need to be changed.
Note: additional minor changes were also applied to the Method Sheet; please refer to the change bars on the right and left text margins for easy change identification.