Health Products

M320 Surgical Microscope (2018-05-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Leica Microsystems (Schweiz) Ag
Max Schmidheiny-Strasse 201
Heerbrugg
9435
SWITZERLAND
M320 Surgical Microscope10448419

Introduction

Affected products

M320 Surgical Microscope

Reason

Three (3) complaints were reported involving mechanical part failures for the Leica M320 Microscope System. These specific systems had failures with the screws that hold the gas spring fixation bracket in place. Due to this failure, the gas spring, which is fixed to the upper part of the parallelogram, cannot hold up the weight of the optics carrier. This may potentially cause the M320 optics carrier to drop to the lower mechanical end stop. This failure has been linked to an identified batch of non-conforming screws.