ABL90 FLEX - Analyzer (2018-05-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Version 3.3 MR1 or lower 
Manufacturer
Radiometer Medical Aps
Akandevej 21
Bronshoj
2700
DENMARK
ABL90 FLEX - Analyzer393-090

Introduction

Affected products

ABL90 FLEX - Analyzer

Reason

Radiometer has recently become aware that the ABL90 FLEX in a very rare specific situation can aspirate CAL 2 solution instead of rinse solution. The specific situation can occur when the following take place:
  • The analyzer is in ready mode.
  • Operator lifts the inlet and then closes it again, thereby initiating a rinse.
  • Immediately after lifts the inlet and quickly closes it again.


This may cause incorrect measurements for the several parameters, showing a high pattern for pH, Na+, Cl− and a low pattern for pCO2, K+ and Ca2+. Some of the parameters will not have an error message or question mark indicating a problem with the result. The wrong reporting will continue until the analyzer flow selector has reset.

Important: The flow selector will reset as part of the automatic error removal process performed during repeated automatic calibrations, QC measurements or during installation of the solution pack.

Risk for the patient, the described error may potentially cause a patient with either:
  • Moderate to severe Hyperkalemia
  • Severe Hypoxia
  • Severe Carbon Monoxide Poisoning