Health Products

RayStation, RayPlan, and InverseArc 1.0 (2018-06-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
4.3.0.14
Manufacturer
RaySearch Laboratories AB (PUBL)
Sveavagen 25
Stockholm
11134
SWEDEN
B. InverseArc 1.04.3.0.14
6.0.0.24
Manufacturer
RaySearch Laboratories AB (PUBL)
Sveavagen 25
Stockholm
11134
SWEDEN
C. RayPlan2
More than 10 numbers, contact manufacturer.
Manufacturer
RaySearch Laboratories AB (PUBL)
Sveavagen 25
Stockholm
11134
SWEDEN
A. RayStation4.0.3

Introduction

Affected products

A. RayStation
B. InverseArc 1.0
C. RayPlan

Reason

When checking the dose distribution for an extreme off-axis field, it was noticed that the dose in the corners did not match that of the previously used planning system. When investigating the beam model, it was found that the beam profile corrections had inappropriate values outside the 20 cm radius.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime