Health Products

VITROS Chemistry Products Na+ Slides (2018-06-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Lots From Gen:8,13,14,16,17,18VITROS Chemistry Products Na+ Slides
Manufacturer
Ortho Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
8379034

Introduction

Affected products

VITROS Chemistry Products Na+ Slides

Reason

Ortho Clinical Diagnostics (Ortho) identified the potential for positively biased results to occur for serum/plasma and urine samples when using the affected VITROS Na+ Slides. The results may be positively biased and quality control fluids may be outside of the expected ranges. The amount of bias in the VITROS Na+ method is dependent upon multiple sources of variability.

Two separate issues have been identified:
  • VITROS Na+ Slides, gens 8 & 13: positively biased urine sample results
  • VITROS Na+ Slides gens 13, 14, 16, 17 & 18: positively biased serum or plasma sample results