Health Products

LIFECODES HLA-B SSO and HLA-B ERES SSO Typing Kit (2018-03-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
01117B
Manufacturer
Immucor GTI Diagnostics, Inc.,
20925 Crossroads Circle,
Waukesha
53186
UNITED STATES
B. LIFECODES HLA-B ERES SSO Typing Kit628917
01117A
Manufacturer
Immucor GTI Diagnostics, Inc.,
20925 Crossroads Circle,
Waukesha
53186
UNITED STATES
A. LIFECODES HLA-B SSO Typing Kit628915

Introduction

Affected products

A. LIFECODES HLA-B SSO Typing Kit
B. LIFECODES HLA-B ERES SSO Typing Kit

Reason

As a result of a complaint investigation, Immucor GTI Diagnostics, Inc. has determined the serological group for the B*27:08 allele was incorrectly assigned as BW4. The correct serological group assignment for B*27:08 is BW6. This assignment was part of the 3.29.0 allele database update.