Health Products

StraightShot Microdebrider System XPS Blades (2018-05-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Medtronic Xomed Inc.
6743 Southpoint Drive North
Jacksonville
32216
Florida
UNITED STATES
StraightShot Microdebrider System XPS Blades1883504

Introduction

Affected products

StraightShot Microdebrider System XPS Blades

Reason

This voluntary recall is being conducted due to potential for irrigation port blockage. The irrigation pin used to manufacture (molding) the hub outer tube was incorrectly placed in the mold which resulted in blockage of the irrigation lumen. Medtronic Puerto Rico Operations Company (MPROC) presented an issue in which the 66d1204 hub outer tube w/barb 3.5mm molded hub manufactured at Medtronic ENT Jacksonville had one cavity with a blocked irrigation port.