Minicon Clinical Sample Concentrators (B-15 & CS-15) (2018-06-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Merck Millipore Ltd.
Tullagreen
Carrigtwohill, Cork
T45 KD29
IRELAND
Minicon Clinical Sample Concentrators (B-15 & CS-15)9031

Introduction

Affected products

Minicon Clinical Sample Concentrators (B-15 & CS-15)

Reason

Since the 22nd of February 2017, 39 Minicon IVD complaints have been received.  97% of the complaints received were for device failing to concentrate the samples within the internal specification of less than or equal to 3 hours. A trend report to the Irish Competent Authority (HPRA) was submitted on March 7th 2018, (HPRA ref v35094).  On May 3rd 2018, the HPRA requested that a field safety notice be issued based on their opinion that there exists a potential risk to patient safety.