Invia Motion Negative Pressure Wound Therapy (NPWT) System (2018-06-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1676710
Manufacturer
Medela AG
Invia Motion Negative Pressure Wound Therapy (NPWT) System087.4012

Introduction

Affected products

Invia Motion Negative Pressure Wound Therapy (NPWT) System

Reason

Medela AG identified an issue where the unforeseen user battery charging behaviour has led to a malfunction. This malfunction was identified only in the case where the user repeatedly plugged and unplugged the pump to the power source and the battery was fully charged. Technically, this can cause the safety fuse to temporarily open the battery connection. When this situation occurs, the device no longer recognises the presence of the battery and notifies the user with a visual and audible "battery missing" notification. In case the device is unplugged in this situation, this will cease operation of the device.