The ABL800 analyzer could potentially induce patient mix-up when performing a query to AQURE, Radiance, or HIS/LIS system and “confirm pre-registered data” is enabled. The problem occurs if "confirmation of demographics" window is open because the user is in the process of updating a sample (e.g. sample #161) with patient demographics requested from HIS/LIS. If the user starts a new measurement (e.g. sample #165) before "confirmation of demographics" window is closed the new sample #165 will incorrectly be updated instead of #161 with the patient demographics shown in the "confirmation of demographics" window. Thus #165 is updated with patient demographics belonging to #161.