Health Products

ABL800 Analyzers (2018-06-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
Radiometer Medical Aps
Akandevej 21
Bronshoj
2700
DENMARK
ABL800 Analyzers393-800

Introduction

Affected products

ABL800 Analyzers

Reason

The ABL800 analyzer could potentially induce patient mix-up when performing a query to AQURE, Radiance, or HIS/LIS system and “confirm pre-registered data” is enabled. The problem occurs if "confirmation of demographics" window is open because the user is in the process of updating a sample (e.g. sample #161) with patient demographics requested from HIS/LIS. If the user starts a new measurement (e.g. sample #165) before "confirmation of demographics" window is closed the new sample #165 will incorrectly be updated instead of #161 with the patient demographics shown in the "confirmation of demographics" window.  Thus #165 is updated with patient demographics belonging to #161.